Header image  

 

PLEASE VIEW THE RELEVANT VIDEO BEFORE ATTENDING THE LAB CLASS

 
line decor
  
line decor
 
 
 


Melting Point: Theory and Practical Lab Tips

Thermometer Calibration: Theory and Practical Lab Tips

Example of a Typical Experiment

FAQ: What's the dial on the Gallenkamp [melting apparatus] instrument for?

FAQ: What is the best way to determine the melting point of my sample?

FAQ: Can I use the boost switch to quickly determine the melting point?

FAQ: When you observe a melting point, you are watching a solid change into a liquid. Do you record the temperature when you see the first drop of liquid and record a second temperature when the sample is a liquid?

FAQ: How do crystal size, sample size, rate of heating, or having a wet sample influence the accuracy of measuring the melting point and/or its range?

FAQ: During lab class, can I leave my thermometer in the Gallenkamp apparatus between melting point measurements? What happens if I leave it in there?

FAQ: My Gallenkamp is not working!

FAQ: What do I do with my old melting point tubes?

FAQ: When taking multiple melting points, must you wait until the Gallenkamp apparatus cools down to room temperature before measuring the melting range of the next sample?

FAQ: Is there any way to quickly cool the Gallenkamp instrument?

FAQ: How much sample should I put into the melting point tube?

FAQ: How come, I can fit only two melting point tubes into the Gallenkamp when there is space for three?

FAQ: Suppose I have two pure compounds, one which melts at 100°C and the other melting at 50°C. Shouldn't the mixed melting point of a 1:1 mixture be depressed for both compounds, i.e., even less than 50°C?

FAQ: What is this calibration all about?

 

Melting Point

Melting point is the temperature at which the solid phase is in equilibrium with the liquid phase.

In practice, however, a melting point is actually recorded as a temperature range: the first temperature, when the first drop of liquid forms; the second temperature, when the last crystal dissolves.

The melting point provides reliable information about the purity and identity of a compound.
A pure solid has a sharp melting point, i.e. a narrow melting range (1-2oC).
The presence of impurities will give a depressed (lowered) melting point with a wide melting range.
A compound is also pure when the mixed melting point of equal portions of the compund and an authentic melting point sample is sharp and not depressed.
The identity of a solid can be determined when pure authentic samples of compounds, (so-called melting point standards) are available for measuring mixed melting points.

Different factors can influence the accuracy of measured melting ranges (crystal size, crystal quantity, rate of heating, dryness of crystals).

Return to Top


Thermometer Calibration

Due to variations in the size of their internal bore, laboratory thermometers differ in accuracy and need to be calibrated.

Melting point standards (known compounds with known, sharp melting points) are used to calibrate a thermometer.

Calibration of a thermometer involves determining the deviation of the measured temperature from a published value for a series of known compounds. This allows you to find the thermometer errors and thus the correction factors.

Two kinds of graph showing the relationship between the temperature a thermometer measures and the actual temperature may be used to correct a temperature measurement:
Observed melting point temperature on the x-axis and the actual (literature) temperature on the y-axis.
Observed melting point temperature on the x-axis and the difference to the actual (literature) temperature on the y-axis.

Return to Top


 A Typical Experiment

Earl N. Meyer realized that his thermometer was inaccurate. He and his partner, Phil Terpaper, decided to "calibrate" it, that is, find the error in temperature reading at specific temperatures.

Phil gathered together the necessary equipment and "melting point standards" (known compounds with known sharp melting point ranges).

He decided to follow the advice of Todd, the Super TA, and measure the temperature ranges of the low melting componds first. After nearly embarrassing himself by trying to fill a melting point tube with ice, Phil inserted his thermometer in a beaker containing ice-water to obtain the 0° C reading. To prevent the thermometer from falling down while in the beaker, he clamped a universal clamp to the monkey bars and secured his thermometer with it.

Phil then packed a melting point tube with ca. 3 mm of finely powdered chemical obtained from the available melting point standard.

Using the Gallenkamp apparatus and controlling the heating rate by adjusting the rheostat, he measured the melting point range for each compound. He noted the temperature when the first drop of liquid appeared and when all of the solid had melted.

From his data, Phil prepared two thermometer calibration graphs:

* First he plotted the midpoint of his observed melting point range versus the published melting point for the known compounds (melting point standards). Using this graph Phil was able to directly interpolate the corrected temperature for a specific observed melting point.
* Secondly, Phil plotted a correction factor [the difference between the observed and published melting point] versus the observed melting point, this time using the upper end of the melting range. With this graph, Phil was able to determine his thermometer error at a given temperature.

Later Phil can use either graph to correct for inaccuracies in his thermometer.

Return to Top

 

Practical Lab Tips

Melting Point

Make sure your sample is dry and crushed to a fine powder.

Pack the solid sample in the bottom of the melting point tube by tapping the tube on the desk.

Insert the closed end of the melting point tube into the Gallenkamp block.

Do not fill too little or too much sample into the melting point tube. Approximately 1-2 mm is right.

If you are unsure about your melting point measurement, prepare another sample and measure the melting temperatures again. Do not reuse your old mp tube, since melting often causes decomposition.

The crystals will shrivel up (simmer) before you will see a liquid drop forming. Do not take the first temperature yet, but wait for the first drop to form before noting the temperature. Likewise, wait for the last crystal to dissolve before noting the second temperature of the melting point range.

Be careful with your thermometer. Try to avoid exposing it to sudden temperature drops. DO NOT place it into ice-water after it has been heated. The thermometer will break! Wait until the thermometer bulb cools to room temperature.

Return to Top

 

Thermometer Calibration

A calibration graph should have:
Title
Axis labels (units of measurements), telling which axis provides which information
Scaling along the axis, i.e., how many lines of the graph correspond to how many units of temperature
Clear marking (dot or cross) of the midpoint of the observed melting point and the literature melting point for each melting point standard
Connect the dots with a straight line

Use one of the two calibration graphs to correct temperature measurements made with this thermometer.

Observed melting point temperature vs. actual (literature) temperature:
Find your measured temperature on the x-axis, then look for the corresponding actual temperature on the y-axis.

Observed melting point temperature vs. difference to the actual (literature) temperature:
Find your measured temperature on the x-axis, then look on the y-axis to determine the correction factor (error size) by which you need to adjust your observed temperature reading.

Return to Top

 

FAQs

Q: What's the dial on the Gallenkamp [melting apparatus] instrument for?

The dial (rheostat) is used to vary the heating rate when measuring a melting temperature.

- A low rheostat setting (e.g., '3') allows for gradual heating, suitable for low melting compounds ( e.g., < 80°C).
- A high setting (e.g., '8') allows for fast heating, suitable for measuring the melting range of higher melting compounds (e.g., > 150°C).

However, each Gallenkamp apparatus is somewhat unique and a rheostat setting of for example '6' on one Gallenkamp may result in a faster heating in comparison to another.

Return to Top

 

Q: What is the best way to determine the melting point of my sample?

If you know the melting point of the compound:
It is a good idea to first heat rapidly at a high rheostat setting to bring the temperature to approximately 15°C below the anticipated melting point. Then turn the rheostat back to a lower setting to gradually heat through the melting range to get accurate temperature measurements.

If you do not know the melting point of the compound:
First obtain an approximate melting point range by rapid heating at a high rheostat setting. Wait for your Gallenkamp instrument to cool approximately 20°C below the anticipated melting point temperature. Then you can proceed as described above.

Return to Top

 

Q: Can I use the boost switch to quickly determine the melting point?

Theoretically, yes. However, this strategy will not be accurate because you may observe a broad melting range and falsely conclude that your compound is impure when, in fact, it is pure. Rapid heating does not allow sufficient time for your compound to melt completely at a given temperature.

If you are running out of time at the end of a lab class, it is a better practice to prepare a melting point tube, and measure the melting point later in the Melting Point Room. Do not forget to take your thermometer with you!

Return to Top

 

Q: When you observe a melting point, you are watching a solid change into a liquid. Do you record the temperature when you see the first drop of liquid and record a second temperature when the sample is a liquid?

Yes. That is the correct technique for measuring since the so-called melting point is actually a melting range.

Return to Top

 

Q: How do crystal size, sample size, rate of heating, or having a wet sample influence the accuracy of measuring the melting point and/or its range?

Well, let's look at some examples. What do you think will happen in each case?

Crystal size:
Imagine you hold a glass of strawberry juice with big ice cubes in your left hand and a glass of strawberry slush in your right hand. Both glasses contain the same total amount of ice. Now, which glass will have ice floating in it for a longer time?

Sample size:
Suppose you have a large and a small icicle. When you heat them up, will they start melting at the same time? Will they be completely melted at the same time? How does the melting range of the two icicles differ?

Rate of heating:
Imagine you have another two icicles, but this time they are of the same size. You heat one icicle slowly (i.e., give it time to melt at a constant temperature) and the other icicle at a rapidly increasing temperature. What will happen to the melting range?

A wet sample:
Some Antifreeze dripped onto one icicle and it is now wet. How will this affect the melting point of ice? Will Antifreeze cause the melting point of ice to be lower or higher?

Return to Top

 

Q: During lab class, can I leave my thermometer in the Gallenkamp apparatus between melting point measurements? What happens if I leave it in there?

When you are measuring the melting point and turn the dial on the Gallenkamp, it is the Gallenkamp apparatus which is heating up. The thermometer only indicates the temperature of the heating block.

If you need to do other things between melting point measurements and want to stop the temperature of the Gallenkamp from increasing, then simply turn it off or set the rheostat to '0'.

Removing the thermometer from the Gallenkamp between measuring melting points will not change the temperature of the apparatus. It is even a good idea to leave the thermometer in the Gallenkamp instrument during your lab class, because the thermometer is secured and less likely to be accidentally broken.

Return to Top

 

Q: My Gallenkamp is not working!

Did you plug it in and turn it on? You can easily verify this by checking that the Gallenkamp lamp is on. If this is not the case, then report the problem to your TA.

Return to Top

 

Q: What do I do with my old melting point tubes?

During the lab class you can store your used melting point tubes in the black storage tube on the right side of the Gallenkamp.

At the end of the lab class, please dispose the used melting point tubes into the yellow waste container labelled "broken glass".

Return to Top

 

Q: When taking multiple melting points, must you wait until the Gallenkamp apparatus cools down to room temperature before measuring the melting range of the next sample?

When measuring the melting points of multiple solids in succession you don't need to wait for the Gallenkamp apparatus [melting point apparatus] to cool to room temperature if you plan your work!

First measure the melting temperatures of lower-melting solids, then measure the melting temperatures of the higher-melting solids.

The Gallenkamp apparatus should be allowed to cool approximately 20°C below the anticipated melting point of the next sample.

Return to Top

 

Q: Is there any way to quickly cool the Gallenkamp instrument?

Yes. Connect the rubber tubing (hose) to the air outlet fixture located on the student bench. Blow a gentle stream of cold air onto the heating block, but do not let the rubber tubing touch the plate. Observe the changing temperature on the thermometer placed in the Gallenkamp.

Return to Top

 

Q: How much sample should I put into the melting point tube?

Fill the melting point tube with enough sample to cover about 2 mm of the tube. Tap the melting point tube a few times onto the desk to ensure that the sample is packed evenly at the closed end of the tube.

Make sure your sample is dry. A wet sample will NOT move easily down the tube, if at all!

Return to Top

 

Q: How come, I can fit only two melting point tubes into the Gallenkamp when there is space for three?

Sometimes a melting point tube breaks inside the Gallenkamp apparatus. Usually you can remove it by inverting the Gallenkamp. However, be careful that you first remove other melting point tubes and the thermometer before turning over the apparatus.

Return to Top

Q: Suppose I have two pure compounds, one which melts at 100°C and the other melting at 50°C. Shouldn't the mixed melting point of a 1:1 mixture be depressed for both compounds, i.e., even less than 50°C?

The typical melting behavior for most 1:1 two-compound mixtures will result in a depression of the melting temperature below that of the individual compounds and a broadening of the melting range. Different molar ratios of the two compounds give different melting temperatures, the lowest melting point of any mixture being the so-called eutectic temperature.

Return to Top

 

Q: What is this calibration all about?

The temperatures you will measure throughout the lab course may differ from the "actual" temperatures. This is because the thermometer you use is relatively cheap and its internal bore is not very uniform.

By measuring the melting ranges of several so-called "melting point standards" (known compounds with sharp, known melting points) you can find the relationship between the temperature your thermometer measures and the actual temperature.

This relationship can be plotted in form of a graph. Two sorts of graphs are possible:

1. You can plot the observed melting point temperature on the x-axis and the actual (literature) temperature on the y-axis. In practice, find your measured temperature on the x-axis, then look for the corresponding actual temperature on the y-axis.
2. Plot the observed melting point temperature on the x-axis and the difference to the actual (literature) temperature on the y-axis. In practice, find your measured temperature on the x-axis, then look on the y-axis to determine the correction factor (error size) by which you need to adjust your observed temperature reading.

In subsequent experiments you can correct your temperature measurements (e.g., m.p. or b.p.) for errors inherent in your thermometer by using the data obtained in your graph. REMEMBER though: Calibration is individual!

Look in the lab manual for further information on how to calibrate, plot, and use the calibration graph.

Return to Top